AED Defibrillator Monthly Check
This procedure defines the monthly inspection, verification, and documentation process for all automated external defibrillators (AEDs) on site. The goal is to ensure every unit is fully operational, within manufacturer expiry dates, and ready for immediate emergency deployment — meeting WorkSafeBC and CSA Z432 compliance requirements.
To ensure all automated external defibrillators (AEDs) on site are fully operational, within manufacturer expiry dates for batteries and electrode pads, and ready for emergency use. This procedure establishes a consistent monthly inspection cadence that meets regulatory requirements under WorkSafeBC OHS Regulation Part 32 — Electrical Equipment and aligns with CSA Z432 — Safeguarding of Machinery and manufacturer maintenance guidelines. Early detection of faults, expired components, or cabinet damage prevents equipment failure during a cardiac emergency.
This procedure applies to every AED unit located within the facility perimeter — including main production floor, warehouse aisles, office corridors, break rooms, remote outbuildings, and vehicle-mounted units (if carried by field crews). It covers visual cabinet inspection, status indicator verification, battery and electrode pad expiry validation, accessory inventory, and documentation logging. This SOP does not cover advanced troubleshooting, internal repairs, or software updates — those must be performed by a qualified AED service technician per manufacturer guidelines.
- Never block, lock, or obstruct AED cabinets during inspections. Ensure the unit is accessible at all times. If a fault is found, tag the cabinet with a clearly visible "Out of Service — Do Not Use" sign and notify the safety manager immediately.
- Wear disposable gloves if handling used electrode pads, exposed adhesive surfaces, or any equipment that may have contacted bodily fluids. Follow site biohazard disposal protocols for contaminated items.
- Lithium batteries must not be punctured, crushed, incinerated, or exposed to water. Dispose of expired or defective batteries according to manufacturer instructions and local hazardous waste regulations.
- Do not open the AED internal casing, disconnect internal wiring, or attempt any repairs. This monthly check is limited to user-serviceable components (battery, pads, cabinet). All other service requires a certified technician.
- Any deficiency — red status light, expired battery or pads, damaged cabinet, missing accessories — must be recorded on a non-conformance report and escalated to the site safety manager within 24 hours. Do not leave a faulty unit in service.
- AED unit with corresponding manufacturer operation manual
- Spare electrode pad set (sealed, within expiry date)
- Spare battery pack (within expiry date, correct model)
- Site-specific AED inspection logbook or digital form
- Pen or mobile device for documentation
- Disposable nitrile gloves (one pair per inspection)
- Small flashlight for visual inspection of cabinet and connectors
- "Inspected — OK" date sticker or tag (if used at your site)